Vulnerability Description
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
CVSS Score
CRITICAL
Affected Products
| Vendor | Product | Versions |
|---|---|---|
| Baxter | Sigma Spectrum Infusion System Firmware | 6.05 |
| Baxter | Sigma Spectrum Infusion System | - |
| Baxter | Wireless Battery Module | 16 |
Related Weaknesses (CWE)
References
- https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01MitigationThird Party AdvisoryUS Government Resource
- https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01MitigationThird Party AdvisoryUS Government Resource
FAQ
What is CVE-2014-5432?
CVE-2014-5432 is a vulnerability with a CVSS score of 9.8 (CRITICAL). Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may b...
How severe is CVE-2014-5432?
CVE-2014-5432 has been rated CRITICAL with a CVSS base score of 9.8/10. This is considered a critical vulnerability requiring immediate attention.
Is there a patch for CVE-2014-5432?
Check the references section above for vendor advisories and patch information. Affected products include: Baxter Sigma Spectrum Infusion System Firmware, Baxter Sigma Spectrum Infusion System, Baxter Wireless Battery Module.