Vulnerability Description
An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
CVSS Score
CRITICAL
Affected Products
| Vendor | Product | Versions |
|---|---|---|
| Baxter | Sigma Spectrum Infusion System Firmware | 6.05 |
| Baxter | Sigma Spectrum Infusion System | - |
| Baxter | Wireless Battery Module | 16 |
Related Weaknesses (CWE)
References
- https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01MitigationThird Party AdvisoryUS Government Resource
- https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01MitigationThird Party AdvisoryUS Government Resource
FAQ
What is CVE-2014-5433?
CVE-2014-5433 is a vulnerability with a CVSS score of 9.8 (CRITICAL). An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700B...
How severe is CVE-2014-5433?
CVE-2014-5433 has been rated CRITICAL with a CVSS base score of 9.8/10. This is considered a critical vulnerability requiring immediate attention.
Is there a patch for CVE-2014-5433?
Check the references section above for vendor advisories and patch information. Affected products include: Baxter Sigma Spectrum Infusion System Firmware, Baxter Sigma Spectrum Infusion System, Baxter Wireless Battery Module.