CRITICAL · 9.8

CVE-2015-3954

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges o...

Vulnerability Description

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

CVSS Score

9.8

CRITICAL

CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Attack Vector
NETWORK
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
NONE
Scope
UNCHANGED
Confidentiality
HIGH
Integrity
HIGH
Availability
HIGH

Affected Products

VendorProductVersions
PifzerPlum A\+ Infusion System Firmware<= 13.4
PifzerPlum A\+ Infusion System-
PifzerPlum A\+3 Infusion System Firmware<= 13.6
PifzerPlum A\+3 Infusion System-
PifzerSymbiq Infusion System Firmware<= 3.13
PifzerSymbiq Infusion System-

Related Weaknesses (CWE)

References

FAQ

What is CVE-2015-3954?

CVE-2015-3954 is a vulnerability with a CVSS score of 9.8 (CRITICAL). Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges o...

How severe is CVE-2015-3954?

CVE-2015-3954 has been rated CRITICAL with a CVSS base score of 9.8/10. This is considered a critical vulnerability requiring immediate attention.

Is there a patch for CVE-2015-3954?

Check the references section above for vendor advisories and patch information. Affected products include: Pifzer Plum A\+ Infusion System Firmware, Pifzer Plum A\+ Infusion System, Pifzer Plum A\+3 Infusion System Firmware, Pifzer Plum A\+3 Infusion System, Pifzer Symbiq Infusion System Firmware.