Vulnerability Description
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
CVSS Score
CRITICAL
Affected Products
| Vendor | Product | Versions |
|---|---|---|
| Pifzer | Plum A\+ Infusion System Firmware | <= 13.4 |
| Pifzer | Plum A\+ Infusion System | - |
| Pifzer | Plum A\+3 Infusion System Firmware | <= 13.6 |
| Pifzer | Plum A\+3 Infusion System | - |
| Pifzer | Symbiq Infusion System Firmware | <= 3.13 |
| Pifzer | Symbiq Infusion System | - |
Related Weaknesses (CWE)
References
- https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01MitigationThird Party AdvisoryUS Government Resource
- https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01MitigationThird Party AdvisoryUS Government Resource
FAQ
What is CVE-2015-3956?
CVE-2015-3956 is a vulnerability with a CVSS score of 9.8 (CRITICAL). Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pum...
How severe is CVE-2015-3956?
CVE-2015-3956 has been rated CRITICAL with a CVSS base score of 9.8/10. This is considered a critical vulnerability requiring immediate attention.
Is there a patch for CVE-2015-3956?
Check the references section above for vendor advisories and patch information. Affected products include: Pifzer Plum A\+ Infusion System Firmware, Pifzer Plum A\+ Infusion System, Pifzer Plum A\+3 Infusion System Firmware, Pifzer Plum A\+3 Infusion System, Pifzer Symbiq Infusion System Firmware.