HIGH · 7.1

CVE-2018-10596

Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading...

Vulnerability Description

Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.

CVSS Score

7.1

HIGH

CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H
Attack Vector
NETWORK
Attack Complexity
HIGH
Privileges Required
LOW
User Interaction
REQUIRED
Scope
UNCHANGED
Confidentiality
HIGH
Integrity
HIGH
Availability
HIGH

Affected Products

VendorProductVersions
Medtronic2090 Carelink Programmer FirmwareAll versions
Medtronic2090 Carelink Programmer-

Related Weaknesses (CWE)

References

FAQ

What is CVE-2018-10596?

CVE-2018-10596 is a vulnerability with a CVSS score of 7.1 (HIGH). Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading...

How severe is CVE-2018-10596?

CVE-2018-10596 has been rated HIGH with a CVSS base score of 7.1/10. Review the CVSS metrics above for detailed severity breakdown.

Is there a patch for CVE-2018-10596?

Check the references section above for vendor advisories and patch information. Affected products include: Medtronic 2090 Carelink Programmer Firmware, Medtronic 2090 Carelink Programmer.