HIGH · 8.8

CVE-2018-18562

An issue was discovered in Roche Accu-Chek Inform II Base Unit / Base Unit Hub before 03.01.04 and CoaguChek / cobas h232 Handheld Base Unit before 03.01.04. Weak access credentials may enable attacke...

Vulnerability Description

An issue was discovered in Roche Accu-Chek Inform II Base Unit / Base Unit Hub before 03.01.04 and CoaguChek / cobas h232 Handheld Base Unit before 03.01.04. Weak access credentials may enable attackers in the adjacent network to gain unauthorized service access via a service interface.

CVSS Score

8.8

HIGH

CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Attack Vector
ADJACENT_NETWORK
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
NONE
Scope
UNCHANGED
Confidentiality
HIGH
Integrity
HIGH
Availability
HIGH

Affected Products

VendorProductVersions
RocheAccu-Chek Inform Ii Firmware< 03.01.04
RocheAccu-Chek Inform Ii-
RocheCobas H 232 Firmware< 03.01.04
RocheCobas H 232-
RocheCoaguchek Firmware< 03.01.04
RocheCoaguchek-
RocheBase Unit Hub Firmware< 03.01.04
RocheBase Unit Hub-

Related Weaknesses (CWE)

References

FAQ

What is CVE-2018-18562?

CVE-2018-18562 is a vulnerability with a CVSS score of 8.8 (HIGH). An issue was discovered in Roche Accu-Chek Inform II Base Unit / Base Unit Hub before 03.01.04 and CoaguChek / cobas h232 Handheld Base Unit before 03.01.04. Weak access credentials may enable attacke...

How severe is CVE-2018-18562?

CVE-2018-18562 has been rated HIGH with a CVSS base score of 8.8/10. Review the CVSS metrics above for detailed severity breakdown.

Is there a patch for CVE-2018-18562?

Check the references section above for vendor advisories and patch information. Affected products include: Roche Accu-Chek Inform Ii Firmware, Roche Accu-Chek Inform Ii, Roche Cobas H 232 Firmware, Roche Cobas H 232, Roche Coaguchek Firmware.