Vulnerability Description
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
CVSS Score
MEDIUM
Related Weaknesses (CWE)
References
- https://static.draeger.com/security/download/2020-06-PSA-20-169-01-Security-Advi
- https://www.vulncheck.com/advisories/dr-ger-perseus-a500-dos-via-medibus-interfa
FAQ
What is CVE-2019-25723?
CVE-2019-25723 is a vulnerability with a CVSS score of 4.0 (MEDIUM). Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted no...
How severe is CVE-2019-25723?
CVE-2019-25723 has been rated MEDIUM with a CVSS base score of 4.0/10. Review the CVSS metrics above for detailed severity breakdown.
Is there a patch for CVE-2019-25723?
Check the references section above for vendor advisories and patch information. Review vendor security bulletins for remediation guidance.