MEDIUM · 4.0

CVE-2019-25723

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted no...

Vulnerability Description

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

CVSS Score

4.0

MEDIUM

CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L
Attack Vector
NETWORK
Attack Complexity
HIGH
Privileges Required
NONE
User Interaction
NONE
Scope
CHANGED
Confidentiality
NONE
Integrity
NONE
Availability
LOW

Related Weaknesses (CWE)

References

FAQ

What is CVE-2019-25723?

CVE-2019-25723 is a vulnerability with a CVSS score of 4.0 (MEDIUM). Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted no...

How severe is CVE-2019-25723?

CVE-2019-25723 has been rated MEDIUM with a CVSS base score of 4.0/10. Review the CVSS metrics above for detailed severity breakdown.

Is there a patch for CVE-2019-25723?

Check the references section above for vendor advisories and patch information. Review vendor security bulletins for remediation guidance.