HIGH · 7.5

CVE-2020-12037

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Pa...

Vulnerability Description

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.

CVSS Score

7.5

HIGH

CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Attack Vector
NETWORK
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
NONE
Scope
UNCHANGED
Confidentiality
HIGH
Integrity
NONE
Availability
NONE

Affected Products

VendorProductVersions
BaxterPrismaflex FirmwareAll versions
BaxterPrismaflex-
BaxterPrismax Firmware< 3.0
BaxterPrismax-

Related Weaknesses (CWE)

References

FAQ

What is CVE-2020-12037?

CVE-2020-12037 is a vulnerability with a CVSS score of 7.5 (HIGH). Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Pa...

How severe is CVE-2020-12037?

CVE-2020-12037 has been rated HIGH with a CVSS base score of 7.5/10. Review the CVSS metrics above for detailed severity breakdown.

Is there a patch for CVE-2020-12037?

Check the references section above for vendor advisories and patch information. Affected products include: Baxter Prismaflex Firmware, Baxter Prismaflex, Baxter Prismax Firmware, Baxter Prismax.