CRITICAL · 9.8

CVE-2020-12047

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-...

Vulnerability Description

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-coded credentials.

CVSS Score

9.8

CRITICAL

CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Attack Vector
NETWORK
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
NONE
Scope
UNCHANGED
Confidentiality
HIGH
Integrity
HIGH
Availability
HIGH

Affected Products

VendorProductVersions
BaxterSigma Spectrum Infusion System Firmware8.0
BaxterSigma Spectrum Infusion System-
BaxterWireless Battery Module17

Related Weaknesses (CWE)

References

FAQ

What is CVE-2020-12047?

CVE-2020-12047 is a vulnerability with a CVSS score of 9.8 (CRITICAL). The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-...

How severe is CVE-2020-12047?

CVE-2020-12047 has been rated CRITICAL with a CVSS base score of 9.8/10. This is considered a critical vulnerability requiring immediate attention.

Is there a patch for CVE-2020-12047?

Check the references section above for vendor advisories and patch information. Affected products include: Baxter Sigma Spectrum Infusion System Firmware, Baxter Sigma Spectrum Infusion System, Baxter Wireless Battery Module.