Vulnerability Description
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-coded credentials.
CVSS Score
CRITICAL
Affected Products
| Vendor | Product | Versions |
|---|---|---|
| Baxter | Sigma Spectrum Infusion System Firmware | 8.0 |
| Baxter | Sigma Spectrum Infusion System | - |
| Baxter | Wireless Battery Module | 17 |
Related Weaknesses (CWE)
References
- https://www.us-cert.gov/ics/advisories/icsma-20-170-04Third Party AdvisoryUS Government Resource
- https://www.us-cert.gov/ics/advisories/icsma-20-170-04Third Party AdvisoryUS Government Resource
FAQ
What is CVE-2020-12047?
CVE-2020-12047 is a vulnerability with a CVSS score of 9.8 (CRITICAL). The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-...
How severe is CVE-2020-12047?
CVE-2020-12047 has been rated CRITICAL with a CVSS base score of 9.8/10. This is considered a critical vulnerability requiring immediate attention.
Is there a patch for CVE-2020-12047?
Check the references section above for vendor advisories and patch information. Affected products include: Baxter Sigma Spectrum Infusion System Firmware, Baxter Sigma Spectrum Infusion System, Baxter Wireless Battery Module.