Vulnerability Description
An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.
CVSS Score
MEDIUM
Affected Products
| Vendor | Product | Versions |
|---|---|---|
| Drtrust | Electrocardiogram Pen Firmware | 2.00.08 |
| Drtrust | Electrocardiogram Pen | 2.00.08 |
References
- https://payatu.com/advisory/lack-of-bluetooth-le-encryption-and-access-control-iThird Party Advisory
- https://payatu.com/advisory/lack-of-bluetooth-le-encryption-and-access-control-iThird Party Advisory
FAQ
What is CVE-2020-15486?
CVE-2020-15486 is a vulnerability with a CVSS score of 6.5 (MEDIUM). An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of t...
How severe is CVE-2020-15486?
CVE-2020-15486 has been rated MEDIUM with a CVSS base score of 6.5/10. Review the CVSS metrics above for detailed severity breakdown.
Is there a patch for CVE-2020-15486?
Check the references section above for vendor advisories and patch information. Affected products include: Drtrust Electrocardiogram Pen Firmware, Drtrust Electrocardiogram Pen.