MEDIUM · 6.8

CVE-2020-7590

A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-202...

Vulnerability Description

A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device.

CVSS Score

6.8

MEDIUM

CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Attack Vector
PHYSICAL
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
NONE
Scope
UNCHANGED
Confidentiality
HIGH
Integrity
HIGH
Availability
HIGH

Affected Products

VendorProductVersions
SiemensDca Vantage Analyzer Firmware< 4.5.0.0
SiemensDca Vantage Analyzer-

Related Weaknesses (CWE)

References

FAQ

What is CVE-2020-7590?

CVE-2020-7590 is a vulnerability with a CVSS score of 6.8 (MEDIUM). A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-202...

How severe is CVE-2020-7590?

CVE-2020-7590 has been rated MEDIUM with a CVSS base score of 6.8/10. Review the CVSS metrics above for detailed severity breakdown.

Is there a patch for CVE-2020-7590?

Check the references section above for vendor advisories and patch information. Affected products include: Siemens Dca Vantage Analyzer Firmware, Siemens Dca Vantage Analyzer.