CRITICAL · 10.0

CVE-2021-33885

An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used i...

Vulnerability Description

An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.

CVSS Score

10.0

CRITICAL

CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:N
Attack Vector
NETWORK
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
NONE
Scope
CHANGED
Confidentiality
HIGH
Integrity
HIGH
Availability
NONE

Affected Products

VendorProductVersions
BbraunSpacecom2< 012u000062
BbraunInfusomat Large Volume Pump 871305U-
BbraunSpacestation 8713142U-

Related Weaknesses (CWE)

References

FAQ

What is CVE-2021-33885?

CVE-2021-33885 is a vulnerability with a CVSS score of 10.0 (CRITICAL). An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used i...

How severe is CVE-2021-33885?

CVE-2021-33885 has been rated CRITICAL with a CVSS base score of 10.0/10. This is considered a critical vulnerability requiring immediate attention.

Is there a patch for CVE-2021-33885?

Check the references section above for vendor advisories and patch information. Affected products include: Bbraun Spacecom2, Bbraun Infusomat Large Volume Pump 871305U, Bbraun Spacestation 8713142U.