Vulnerability Description
An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.
CVSS Score
HIGH
Affected Products
| Vendor | Product | Versions |
|---|---|---|
| Bbraun | Spacecom2 | < 012u000062 |
| Bbraun | Infusomat Large Volume Pump 871305U | - |
| Bbraun | Spacestation 8713142U | - |
Related Weaknesses (CWE)
References
- https://www.bbraunusa.com/en.htmBroken Link
- https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-ExploitThird Party Advisory
- https://www.bbraunusa.com/en.htmBroken Link
- https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-ExploitThird Party Advisory
FAQ
What is CVE-2021-33886?
CVE-2021-33886 is a vulnerability with a CVSS score of 8.1 (HIGH). An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external stri...
How severe is CVE-2021-33886?
CVE-2021-33886 has been rated HIGH with a CVSS base score of 8.1/10. Review the CVSS metrics above for detailed severity breakdown.
Is there a patch for CVE-2021-33886?
Check the references section above for vendor advisories and patch information. Affected products include: Bbraun Spacecom2, Bbraun Infusomat Large Volume Pump 871305U, Bbraun Spacestation 8713142U.