Vulnerability Description
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
CVSS Score
MEDIUM
Affected Products
| Vendor | Product | Versions |
|---|---|---|
| Bbraun | Battery-Pack Sp With Wifi Firmware | <= 053l000092 |
| Bbraun | Battery-Pack Sp With Wifi | - |
Related Weaknesses (CWE)
References
- https://www.bbraun.com/productsecurityProduct
- https://www.bbraunusa.com/productsecurityProduct
- https://www.bbraun.com/productsecurityProduct
- https://www.bbraunusa.com/productsecurityProduct
FAQ
What is CVE-2023-0888?
CVE-2023-0888 is a vulnerability with a CVSS score of 4.9 (MEDIUM). An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access t...
How severe is CVE-2023-0888?
CVE-2023-0888 has been rated MEDIUM with a CVSS base score of 4.9/10. Review the CVSS metrics above for detailed severity breakdown.
Is there a patch for CVE-2023-0888?
Check the references section above for vendor advisories and patch information. Affected products include: Bbraun Battery-Pack Sp With Wifi Firmware, Bbraun Battery-Pack Sp With Wifi.