CVE-2024-48966

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain...

Vulnerability Description

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

CVSS Score

10.0

CRITICAL

CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Attack Vector
NETWORK
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
NONE
Scope
CHANGED
Confidentiality
HIGH
Integrity
HIGH
Availability
HIGH

Related Weaknesses (CWE)

References

FAQ

What is CVE-2024-48966?

CVE-2024-48966 is a vulnerability with a CVSS score of 10.0 (CRITICAL). The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain...

How severe is CVE-2024-48966?

CVE-2024-48966 has been rated CRITICAL with a CVSS base score of 10.0/10. This is considered a critical vulnerability requiring immediate attention.

Is there a patch for CVE-2024-48966?