CRITICAL · 9.3

CVE-2024-9832

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthori...

Vulnerability Description

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

CVSS Score

9.3

CRITICAL

CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Attack Vector
LOCAL
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
NONE
Scope
CHANGED
Confidentiality
HIGH
Integrity
HIGH
Availability
HIGH

Related Weaknesses (CWE)

References

FAQ

What is CVE-2024-9832?

CVE-2024-9832 is a vulnerability with a CVSS score of 9.3 (CRITICAL). There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthori...

How severe is CVE-2024-9832?

CVE-2024-9832 has been rated CRITICAL with a CVSS base score of 9.3/10. This is considered a critical vulnerability requiring immediate attention.

Is there a patch for CVE-2024-9832?

Check the references section above for vendor advisories and patch information. Review vendor security bulletins for remediation guidance.