Vulnerability Description
An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.
CVSS Score
CRITICAL
Affected Products
| Vendor | Product | Versions |
|---|---|---|
| Thermofisher | Ion Torrent Onetouch 2 Firmware | - |
| Thermofisher | Ion Torrent Onetouch 2 | - |
Related Weaknesses (CWE)
References
- https://assets.thermofisher.com/TFS-Assets/LSG/manuals/MAN0014388_IonOneTouch2SyProduct
- https://documents.thermofisher.com/TFS-Assets/CORP/Product-Guides/Ion_OneTouch_2Vendor Advisory
- https://www.thermofisher.com/order/catalog/product/4474779Product
FAQ
What is CVE-2025-54304?
CVE-2025-54304 is a vulnerability with a CVSS score of 9.8 (CRITICAL). An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and...
How severe is CVE-2025-54304?
CVE-2025-54304 has been rated CRITICAL with a CVSS base score of 9.8/10. This is considered a critical vulnerability requiring immediate attention.
Is there a patch for CVE-2025-54304?
Check the references section above for vendor advisories and patch information. Affected products include: Thermofisher Ion Torrent Onetouch 2 Firmware, Thermofisher Ion Torrent Onetouch 2.