Vulnerability Description
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.
CVSS Score
CRITICAL
Related Weaknesses (CWE)
References
- https://spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sent
- https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Secu
- https://www.vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthentic
- https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Secu
FAQ
What is CVE-2026-0611?
CVE-2026-0611 is a vulnerability with a CVSS score of 9.8 (CRITICAL). Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed...
How severe is CVE-2026-0611?
CVE-2026-0611 has been rated CRITICAL with a CVSS base score of 9.8/10. This is considered a critical vulnerability requiring immediate attention.
Is there a patch for CVE-2026-0611?
Check the references section above for vendor advisories and patch information. Review vendor security bulletins for remediation guidance.